FSI Customer Service Bulletin

 

Recently FSI has determined a certain sequence of events that can cause damage to the internal calibration constants of the 3D-ACM. This can only occur if power is applied to the 3D-ACM incorrectly.

The problem occurs when the internal battery is disconnected and the operator connects the 3D-ACM Test Cable to the instrument with the external power "live". This causes an internal transient, which may damage the internal calibration constants. You can avoid this problem by connecting the 3D-ACM Test Cable to the end cap and then turn on the external power. If the instrument battery is connected the problem will not occur.

FSI realizes this problem may occur, so we have created a Field Service Kit to prevent the problem from occurring. Contact the Customer Service Department to receive a Field Service Kit.

Sincerely,

John Duchesney

Eng. Mgr., FSI

Instructions for Adding Kick Start Protection

  1. Remove the 3D-ACM from the frame (refer to the 3D-ACM Manual for details).
  2. Remove the Top End Cap from the housing.
  3. Locate the CPU PC Board.
  4. Remove J3 (2 Pin Molex) connector from the CPU PC Board.
  5. Insert this 2 Pin connector into the Field Kit Male connector.
  6. Plug the Field Kit Female 2 Pin Molex connector into J3 on the CPU PC Board.
  7. Insert the Top End Cap into the housing and tighten the End Cap Clamp.
  8. Insert the 3D-ACM into the frame (refer to the 3D-ACM Manual for details).

Your 3D-ACM is ready for deployment.

 